• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #3 L-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #3 L-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5510F301
Medical Device Problem Codes Device Damaged Prior to Use (2284); Packaging Problem (3007)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/13/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
The nurse in the room opened the outside box to open the implant and the inside package was damaged (cracked).
 
Additional Manufacturer Narrative
An event regarding packaging damage involving a triathlon femoral component was reported.The event was confirmed.Based on the visual inspection of the returned packaging it appears that this component packaging was subjected to excessive/inappropriate handling whereby the carton may have been compressed and/or dropped from a height causing damage to the carton and outer blister.Medical records received and evaluation: not performed as the event is related to a packaging issue and no adverse consequences to the patient were reported.-device history review: dhr review for the reported lot determined that the device was manufactured and packed to specification.Complaint history review: there have been no other similar events for the lot referenced.The investigation concluded that the packaging damage was most likely caused by excessive/incorrect handling prior to the opening packaging whereby the carton may have been compressed and/or dropped from a height causing the damage to the carton making it difficult to remove the device.No further investigation for this event is required at this time.
 
Event or Problem Description
The nurse in the room opened the outside box to open the implant and the inside package was damaged (cracked).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #3 L-CEM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key6239321
Report Number0002249697-2017-00160
Device Sequence Number938516
Product Code MBH
UDI-Device Identifier07613327040258
UDI-Public(01)07613327040258(11)160826(17)210831(10)BEH2L
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Catalogue Number5510F301
Device Lot NumberBEH2L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 12/13/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/10/2017
Supplement Report FDA Received Date03/08/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age59 YR
Patient Weight62
-
-