• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/14/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt00 intraocular lens (iol) was explanted due to post refractive issue or over minus issue.Lens was replaced with the same model lens, a 18.5 diopter iol.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 1/17/2017.Device returned to manufacturer ¿ yes.Device evaluation: the device was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this order number to date.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6239935
Report Number9614546-2017-00020
Device Sequence Number16250029
Product Code POE
UDI-Device Identifier05050474605190
UDI-Public(01)05050474605190(17)210922
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/22/2021
Device Model NumberZXT150
Device Catalogue NumberZXT150U195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 12/15/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/10/2017
Supplement Report FDA Received Date03/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-