MAUDE Adverse Event Report: COOK INC UNIVERSA SOFT URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA SOFT URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/27/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event or Problem Description
It was reported by the user facility that the tethered suture stretched and frayed.The physician was able to cut the suture and remove the suture.The new suture was used to restitch the device with no further complications to the patient.The patient did not require any additional procedures nor experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation.A review of the device history record, complaint history, quality controls, specification and visual inspection was conducted on the returned device during the investigation.Incomplete device was returned for investigation.It was discovered that only a tether was returned, which was tangled together.During investigation, the tether was untangled and it was observed that the tether was separated into three segments.Two segments had severely stretched appearances indicating that it was pulled by force stressing the material to the point of snapping to separation.Third segment was noted to have the knot intact.There is no indication that a design or process related failure mode contributed to this event.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.In addition review of device master record did not observe any nonconformance that may have contributed to this incident.Based on the provided information a definitive root cause cannot be established or reported at this time.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA SOFT URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6240305
Report Number1820334-2017-00100
Device Sequence Number755905
Product Code FAD
UDI-Device Identifier10827002233772
UDI-Public(01)10827002233772(17)190824(10)7222603
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K151051
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberUSH-628-RT1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/05/2017
Supplement Date Received by Manufacturer06/20/2017
Initial Report FDA Received Date01/11/2017
Supplement Report FDA Received Date07/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-