| Catalog Number |
050300000 |
| Medical Device Problem Codes |
Crack (1135); Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/08/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Sorin group (b)(4) manufactures the revolution centrifugal blood pump.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The damaged device has been requested for return to sorin group (b)(4) for investigation.A follow-up report will be sent when the investigation is completed.
|
| |
|
Event or Problem Description
|
|
Sorin group (b)(6) received a report that the port of the revolution centrifugal blood pump cracked when the perfusionist attempted to attach the tubing to the centrifugal head outlet during priming.The unit was swapped out for the procedure.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was not returned for investigation.(b)(4).Sorin group (b)(4) manufactures the revolution centrifugal blood pump.The incident occurred in (b)(6).Per exemption number e2016005, sorin group (b)(4) is submitting the report for both sorin (b)(4) (manufacturer) and (b)(4) (importer).Though the involved device was initially thought to be available for return, this was not the case.Therefore, the issue was evaluated based on previously investigated cases for similar issues.Based on these investigations, sorin group (b)(4) believes that the crack may be ascribed to rough handling during transportation, in combination with residual material stress from processing operations.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The product was released in compliance with the specifications.Device not available for return.
|
| |
|
Search Alerts/Recalls
|