| Model Number |
MICL12.1 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Event problem and evaluation codes: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter stated the surgeon implanted a 12.1mm micl12.1 implantable collamer lens, -6.0 diopter, in the patient's left eye (os) on (b)(6) 2016.The lens was explanted on (b)(6) 2016 due to myopic refraction.The lens was exchanged for another same model (12.1mm) but different diopter power (-7.5).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the lens was returned in a lens vial with liquid.A visual inspection of the returned product found a tear on the haptic and a small piece of the haptic missing.(b)(4).
|
| |
|