PHILIPS MEDICAL SYSTEMS HEARTSTART MRX -EMS DEFIBRILLATOR Back to Search Results
Model Number M3536A
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
It was reported to philips that the device is having issues acquiring ecg after an fco was implemented.There was no reported patient involvement/adverse patient impact.
 
Manufacturer Narrative
One processor pca was returned for failure analysis.The reported problem could not be verified or duplicated during lab tests.No fault was found with this processor board.Philips cannot rule out a malfunction of the processor pca at the time of the reported event.The cause was not determined.The available information from this report does not support that this malfunction represents a systemic, design, or labeling problem.No further investigation or action is warranted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX -EMS DEFIBRILLATOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
richa shah
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key6244588
MDR Text Key64934154
Report Number1218950-2017-00306
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/13/2016
Initial Date FDA Received01/11/2017
Supplement Dates Manufacturer ReceivedNot provided
12/13/2016
Supplement Dates FDA Received05/16/2017
08/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/30/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-