ZYNO MEDICAL LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number A2-80071-DF
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/15/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The user-reported leakage issue was confirmed.Affected iv set are under recall #74892.The complaint was determined as not mdr reportable at the initial determination by following company procedures.This mdr is retrospectively filed as a corrective action to address one of the fda inspection observations made on 08/11/2016.
 
Event or Problem Description
The customer reported that "i wanted to alert you to another incident with a chemo spill related to the filter tubing.This one was a little more significant in size, costlier, and caused real distress to the patient and her spouse." the leakage amount was 40-50cc of the medication.No patient injury or harm was involved in this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
ZYNO MEDICAL LLC
177 pine street
natick MA 01760
MDR Report Key6247657
Report Number3006575795-2016-00141
Device Sequence Number13717485
Product Code FPA
UDI-Device Identifier00814371020075
UDI-Public00814371020075
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K132841
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 01/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2017
Device Model NumberA2-80071-DF
Device Lot Number1404003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2015
Initial Report FDA Received Date01/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-