• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZYNO MEDICAL LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZYNO MEDICAL LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number A2-80072
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 04/23/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The user-reported issue was confirmed that back check valve on primary set did not stop secondary drug from backing into primary drip chamber.A supplier corrective action request (scar-002) had been issued to address this issue.The complaint was determined as not mdr reportable at the initial determination by following company procedures.This mdr is retrospectively filed as a corrective action to address one of the fda inspection observations made on (b)(6) 2016.Zyno medical submitted an e-mdr (3006575795-2016-00111_complaint (b)(4)) on 10/26/2016 through the fda's website and received the receipt.But due to some technical difficulties and confusion, the file didn't pass with all three acknowledgements.This is a re-submission regarding the same event.
 
Event or Problem Description
The customer stated that the tubing had fluid backing up into secondary (tubing).No patient was involved in this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
ZYNO MEDICAL LLC
177 pine street
natick MA 01760
MDR Report Key6247796
Report Number3006575795-2016-00111
Device Sequence Number19023386
Product Code FPA
UDI-Device Identifier00814371020112
UDI-Public00814371020112
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K132841
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 01/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA2-80072
Device Lot Number58159KS
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-