• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. EPTFE VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. EPTFE VASCULAR GRAFT Back to Search Results
Catalog Number F7008TWS
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number of the device was provided.The device history records are currently under review.The device is not available for return.The investigation is currently under way.
 
Event or Problem Description
It was reported that during preparation, upon removal of the beading from the vascular graft, the graft allegedly tore.There was no reported patient contact.
 
Additional Manufacturer Narrative
Manufacturing review: device history record (dhr) for lot number vtaq0009 was reviewed and the lot met all release criteria.The dhr review showed that all process parameters were within specification and that this lot of grafts passed all physical testing requirements.No changes have been made to the manufacturing process that would impact this complaint.Visual/microscopic inspection: as the device was not returned, an inspection could not be performed.Functional/performance evaluation: as the device was not returned, an evaluation could not be performed.Medical records review: as medical records were not provided, a review could not be performed.Image/photo review: as medical images were not provided, a review could not be performed.Conclusion: images and medical records were not provided.The investigation was inconclusive for the alleged graft tears.The definitive root cause could not be determined based upon available information.It was unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: do not remove the external spiral support (beading) from any centerflex¿ graft or endflex¿ graft.Attempts to remove the beading may damage the graft wall.If damage occurs, discard the graft.Precautions: when removing the external spiral support (beading) of impra flex¿ grafts, the beading must be removed slowly and at a 90° angle to the graft.Rapid unwinding and/or removal at less than a 90° angle may result in graft damage.Do not use surgical blades or sharp, pointed instruments to remove the beading as this may damage the graft wall.If damage occurs, that segment of the graft should not be used.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during preparation, upon removal of the beading from the vascular graft, the graft allegedly tore.There was no reported patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPTFE VASCULAR GRAFT
Common Device Name
VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6248094
Report Number2020394-2016-01264
Device Sequence Number16376024
Product Code DYF
UDI-Device Identifier00801741023279
UDI-Public(01)00801741023279(17)210228(10)VTAQ0009
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K983064
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 12/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue NumberF7008TWS
Device Lot NumberVTAQ0009
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 12/20/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/12/2017
Supplement Report FDA Received Date01/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-