|
Device is a combination product.(b)(4).Device evaluated by mfr: promus premier,us, mr 3.00x12mm stent delivery system (sds) was returned for analysis.A visual examination of the crimped stent found the entire stent was damaged.The stent was distorted and bunched in a proximal direction; partially covering the proximal marker band and exposing the distal side of the balloon wall for approximately 7mm.Crimp markings were evident on the 7mm of exposed balloon wall, indicating crimp contact between the coated stent and balloon.The balloon cones were reviewed and no issues were noted with the distal balloon cone; the balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.Damage was noted to the proximal balloon cone, it appeared bunched and distorted.The bumper tip of the device showed slight signs of damage to the distal edge of the tip.A visual and tactile examination found multiple hypotube kinks.A visual and tactile examination of the outer and the inner lumen, and mid-shaft revealed a kink within the mid-shaft section, approximately 3mm proximal to the port well.The bi-component bond showed no signs of damage or strain.A 0.014'' guidewire was tracked through the inner lumen, entering via the port weld and exiting via the tip, no tracking difficulties were noted, indicating no issue with or blockages within the inner lumen.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable cause of the reported difficulties may be due interaction with another device.(b)(4).
|
|
Reportable based on analysis completed on 19-dec-2016.It was reported advancing difficulties were encountered and shaft kink occurred.A 3.00x12mm promus premier¿ drug eluting stent was selected to treat the lesion.During preparation, the physician attempted to insert the device through a non-bsc guide catheter, however, the non-bsc guide catheter was kinked and the stent delivery system (sds) was unable to pass through.Upon removal of the sds, it was noticed that the shaft was kinked at the wire exit port.No patient complications were reported.However, returned device analysis revealed stent damage and stent moved on balloon.
|