Model Number 4415
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 12/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint is related to (b)(4)/ medwatch # 1828100-2016-00781.
 
Event Description
It was reported that during the use of the device for a non-clinical activity, the unit was failing electrical safety test.There was no patient involvement.The issue was discovered when the unit was repaired in the field.
 
Manufacturer Narrative
The field service representative (fsr) verified the issue.After allowing the unit to dry out he fully tested the heater cooler (tcm) and it was working fine.The unit operated to manufacturer¿s specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
eileen dorsey
6200 jackson road
ann arbor, MI 48103
7347416074
MDR Report Key6251728
MDR Text Key65100608
Report Number1828100-2017-00020
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/1999
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-