| Catalog Number |
EUP2515X |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Intimal Dissection (1333)
|
| Date of Event |
12/17/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: numerous kinks and stretching were visible on the distal shaft.The device returned deflated with blood and residue present inside the inflation lumen and balloon.Negative prep did not detect a leak possibly due to the residue in the inflation lumen.The device was placed in a waterbath to disperse the residue.On removal of the device negative prep confirmed the presence of a leak.On pressurization of the device a leak was detected; a short radial tear was visible on the working length of the balloon.The balloon material was jagged and uneven along the leak site.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the physician was attempting to use a euphora rx balloon to treat a moderately calcified and slightly tortuous proximal lad lesion.No damage noted to device packaging and no issues were noted when removing the device from the hoop/tray.The device was inspected and negative prep performed with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent.No resistance was encountered when advancing the device to the lesion.It was reported that the device was inflated in the rca once with no issues.It was then delivered to the lad and inflated once.The balloon started losing pressure at 13 atms.The device was not moved or re-positioned prior to the burst.After the case, when the physician removed and attempted to re-inflate the balloon outside the body.He noticed a pin point hole that saline shot out of with substantial pressure.It was not a longitudinal split.The balloon caused a dissection.The physician put in two stents to cover.Patient is doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|