| Model Number |
H7493808012370 |
| Medical Device Problem Codes |
Break (1069); Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.75mm x 12mm quantum¿ maverick¿ balloon catheter was selected for use.However, while the device was still outside the patient's body, the shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product consisted of a quantum maverick balloon catheter.The balloon was tightly folded.There was contrast in the inflation lumen.There was blood in between balloon folds.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube shaft was completely separated 56.5 cm from the hub.The fracture faces were oval as if kinked prior to separation.There were numerous hypotube kinks.There was no evidence of any material or manufacturing deficiencies contributing to the damage.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.75mm x 12mm quantum¿ maverick¿ balloon catheter was selected for use.However, while the device was still outside the patient's body, the shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|