• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Submitted: (b)(6) 2017.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.The pump has been returned and evaluated by product analysis on 12/20/2016 with the following findings.On investigation, the battery compartment was cracked and the display was dim/faded and discolored.
 
Event Description
The pump was returned for investigation.Investigation revealed battery compartment and display screen issue.This report is made based on results of investigation completed on (b)(6) 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6255169
MDR Text Key65252414
Report Number2531779-2017-01291
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial
Report Date 12/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2016
Was the Report Sent to FDA? Yes
Date Manufacturer Received12/20/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-