| Model Number |
H7493790950150 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/19/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older. (b)(4).Device evaluated by mfr: the device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 95% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending (lad) artery.A 5.0mm x15mm maverick¿ xl balloon catheter was advanced to dilate the lesion.However, during inflation, the balloon ruptured.The device was completely removed from the patient's body.The procedure was completed with a different device.No patient complications were reported and the patient's status was good.
|
| |
|