|
|
| Model Number |
BL5111 |
| Medical Device Problem Code |
Particulates (1451)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/15/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sterilization and lot history records were reviewed and found to be acceptable.Report 9 of 16.See 1920664-2017-00012, 00020, 00021, 00022, 00023, 00024, 00025, 00026, 00028, 00029, 00030, 00031, 00032, 00033, 00034.
|
| |
|
Event or Problem Description
|
|
The surgeon has been seeing glimmering specks while irrigating the eye, and is not able to remove the particles.They are not sure what is causing the glimmer.Additional information: there have been no adverse effects, but the surgeon did confirm that he sees the ¿shimmer¿ indefinitely post-op.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the results of our root cause investigation, it has been determined that the reported particulate is not related to any damage or malfunction of the product.Our investigation has concluded that the observed particles are originating from the metallic cap assembled to the needle wrench which when packaged in the same pouch with the silicone irrigation sleeve results in microscopic particles being transferred from the needle wrench cap to the irrigation sleeve.All products have been found to comply with specifications and function as intended.Additionally, there have been no reports of post-surgical adverse reaction.However, due to increased sightings of microscopic particles and our ongoing continuous improvement activities, we have taken measures to identify an internal cleaning process related to needle wrench assembly to address the reported events.Validation activities have been completed and the process has been implemented.In addition, we are reviewing our component specifications related to particulate matter and will revise specifications, where necessary to better align with customer expectations.As a result of this additional information, it has been determined that the issue documented in this report does not meet the criteria of a reportable malfunction and the complaint record will be revised accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|