| Model Number |
M0044500THK20 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/01/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr: unit returned in a generic plastic bag, overall visual revision did not identify failures or evidence that could be lost due to decontamination process.The returned device matches with upn and lot provided by the customer.The device has a kink at the distal section while in the neutral position at 5 cm from the tip.In addition the device has a char in the tip.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable upon analysis completed on 22dec2016.It was reported that the device was unable to ablate.A 7 / 110 / 2.5 / 8-8 qd k2 htd blazer ii xp was selected to treat the lesion.During the procedure, it was noted that the device was unable to ablate.The temperature was reading at 130 and was unable to deliver power.Physician reset system and changed cables but it did not change.The device was then removed and the procedure was completed with a different device.No patient complications were reported.However, device analysis revealed that a char on the distal tip was noted.
|
| |
|