• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number N/A
Medical Device Problem Codes Kinked (1339); Difficult to Remove (1528)
Health Effect - Clinical Codes Awareness during Anaesthesia (1707); Pain (1994)
Date of Event 12/13/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event or Problem Description
It was reported by the user facility that a patient underwent an unspecified procedure with stent placement for well drainage.During an implant removal a few weeks post implant, the stent was not able to come out and was causing the patient pain.The physician preformed scans on the patient but found no problems.The patient was put under anesthesia for the removal of the stent.Once removed it was noted that the stent was kinked.It was reported that the patient post procedure was doing well.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, drawings, device history record, documentation, manufacturing instructions and trends of the device was conducted during the investigation.The complaint device was not returned therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.Review of device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, it is possible that user technique contributed to this event; however, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA FIRM URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6261308
Report Number1820334-2017-00130
Device Sequence Number11001724
Product Code FAD
UDI-Device Identifier10827002233130
UDI-Public(01)10827002233130(17)191020(10)7374455
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K961446
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberN/A
Device Catalogue NumberUFH-600-RT1
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/12/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/18/2017
Supplement Report FDA Received Date03/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-