| Model Number |
2200P-47E |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Fracture (1260); Device Inoperable (1663); Low impedance (2285); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Therapeutic Response, Decreased (2271)
|
| Date of Event |
08/12/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A subject participating in the recharge clinical trial (subject id (b)(6) underwent initial implantation of the maestro rechargeable system on (b)(6) 2011.Subject presented with red light (b)(6) 2015.Interrogation with clinician programmer indicated posterior tip to ring short since (b)(6) 2015 with display of error message.Therapy was restarted by site.Same issue occurred (b)(6) 2015 and again on (b)(6) 2015 with therapy restarted after each event.Maestro rechargeable system was explanted on (b)(6) 2015.Subject experienced seroma following explant which resolved without surgical or medical intervention.
|
| |
|