Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure, Delayed, Uncontrolled (1936); No Code Available (3191)
Date of Event 12/01/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter stated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -8.5 diopter, in the patient's left eye (os) on (b)(6) 2016.The lens was explanted on (b)(6) 2016 due to extreme vaulting, which caused pressure.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Additional data: patient complained of pain and after examination, extreme vaulting was found causing high pressure.Lens was explanted and a lens of shorter length was implanted.The patient is doing well and bcva was 20/20 post second icl implant.(b)(4).
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key6265479
Report Number2023826-2017-00095
Device Sequence Number19067212
Product Code MTA
UDI-Device Identifier00841542103268
UDI-Public(01)00841542103268(17)171130
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/30/2017
Device Model NumberMICL13.2
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 12/21/2016
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date01/19/2017
Supplement Report FDA Received Date02/14/2017
03/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL AND LOT NUMBER UNK; INJECTOR MODEL AND LOT NUMBER UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient Weight43
-
-