| Model Number |
3058 |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Unintended Collision (1429); Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Operates Differently Than Expected (2913); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a patient via a company representative (rep).It was reported that patient followed up in office and reporting that system was not working.There were abnormal impedances noted during interrogation by office staff.They were unable to obtain battery life status during interrogation.Patient reported that she had fallen a couple times and fell on unit.The device had not worked well since.The implantable nuerostimulator (ins) battery was also very loose in the pocket.The full system replacement occurred on (b)(6) 2017.The issue was resolved at time of the report.Patient status was noted as alive, no injury.The indications for use for this patient were urinary dysfunction/sacral nerve stim and gastrointestinal/pelvic floor.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|