Catalog Number 430-98-052
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/29/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the cup pushing on the patient's bladder.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the hemispherical shell was pushing on the patient's bladder.The in-vivo length of patient service for the implant was 2.1 months.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient's hemispherical shell was pushing on the patient's bladder.The surgeon performed this revision to remedy the patient's condition.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP HIP
Common Device Name
FMP HEMISPHERICAL SHELLS W/SCREW HOLES, 52MM, W/P2 COATING
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key6266685
Report Number1644408-2017-00030
Device Sequence Number10993310
Product Code LPH
UDI-Device Identifier00888912084024
UDI-Public(01)00888912084024
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K974093
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/24/2022
Device Catalogue Number430-98-052
Device Lot Number587G1213
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 12/29/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/20/2017
Supplement Report FDA Received Date03/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
010-55-030,LOT 007A1178; 497-36-000,LOT 638C1200; 931-36-752,LOT 761N1950
Outcome Attributed to Adverse Event Required Intervention;
Patient Age89 YR
-
-