| Model Number |
LS3112 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/21/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).One ls3112 ligasure device was received for evaluation, and the reported issue with detection was not confirmed.Testing of the device found it was recognized by the generator and there were no issues with the cord dot-recognition fixture.Sealing was performed multiple times on simulated tissue, and all seal cycles were completed satisfactorily with end tones.However, an alternate and non-contributing issue was identified.Visual inspection found one of the snaps on the jaw was broken and missing.Review of the device history records for this lot found no potentially contributing factors.A 100% visual inspection of the snap pins is performed during production, and it is unlikely the product was damaged when it left the manufacturing facility.Snap damage can be caused by misaligning the snaps during assembly by the user or by multiple assembly attempts.The investigation identified the root cause of the damage to be user error.The instructions for use included with this product lists measures for proper assembly.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a procedure, the device was not recognized when connected to a forcetriad generator.A new device was connected to the same generator and worked well.Examination of the received device on january 6th found one of the snap pins from the jaw is broken and missing.The customer confirmed the piece did not fall within the patient.
|
| |
|