SYNTHES OBERDORF CABLE 4M TO CONSOLE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.021VET |
Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/06/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during an unspecified veterinary surgical procedure, it was observed that the cable to console device had an intermittent activity; and that it had a possible broken wire.There was a five minute delay to the planned surgical procedure.A spare identical device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
During subsequent follow-up with the customer, it was mentioned that the device was in use with an electric pen drive during the event.The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device failed the test adapter test (faulty), there was corrosion on connector ends and there was a displaced pin on socket connector ( pushed in).Therefore, the reported condition was confirmed.The assignable root cause was determined to be due improper cleaning methods and improper handling, which are both classified as user error.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|