Model Number 375-545-000
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 11/15/2016
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to stryker sustainability solutions for evaluation.It was stated the device was disposed of by the customer facility.As the device was not returned for evaluation, visual and functional inspection was unable to be performed.A review of the dhr could not be performed as the lot number and/or serial number was not reported.The reported event could be attributed to the user applying too much force to the device or the arthroscopic shaver may have come into contact with staples, clips or another metal object, resulting in damage to the blade.The instructions for use (ifu) state: - before beginning the procedure, verify compatibility of all instruments and accessories.- plug in and set up the generator according to the instructions in the manufacturers¿ manual.- select an arthroscopic shaver with size, blade, and function most appropriate for the procedure.- inspect the instrument for overall condition and physical integrity.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.- careful handling of the instrument is necessary to avoid damage or breakage as a result of excessive force.- do not apply excessive pressure or ¿side-load¿ the blade during use.Side-loading does not improve the performance of the instrument, can dull the blade, and/or produce metal particulates.- do not allow the arthroscopic shaver to come into contact with staples, clips or any metal object to avoid damage to the blade and possible patient injury.- the tip of the bur or cutter must be irrigated periodically (general recommendation: once a minute) to cool the blade and prevent excised tissues from accumulating.- do not run the instrument without appropriate suction for the duration of the process.Should the device become available for return, the investigation will be reopened at that time.The reported event will continue to be monitored through post-market surveillance.(b)(4).
 
Event Description
It was reported that pieces of the arthroscopic shaver broke off into the patient.The doctor was able to recover all of the pieces.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer Contact
moira barton varty
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key6271235
MDR Text Key65913147
Report Number0001056128-2017-00018
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00885825014223
UDI-Public(01)00885825014223
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K012635
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number375-545-000
Device Catalogue Number375-545-000RR
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-