Model Number ZXT300
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/05/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was scratched during insertion into the eye.Patient contact was reported.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 02/09/2017.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection at 12x microscope magnification showed that part of the haptic was missing.Scratches were seen on surface of the optic.The customer's reported complaint for lens scratched was verified, however, the condition of the return sample do not suggest that the damage was introduced during manufacturing.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue was verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6274965
Report Number9614546-2017-00047
Device Sequence Number734747
Product Code POE
UDI-Device Identifier05050474606357
UDI-Public(01)05050474606357(17)210823
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/23/2021
Device Model NumberZXT300
Device Catalogue NumberZXT300U185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/05/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/24/2017
Supplement Report FDA Received Date04/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-