• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE GUNTHER TULIP JUGULAR & FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GUNTHER TULIP JUGULAR & FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-UNI
Device Problems Positioning Failure (1158); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/10/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Similar to device under 510(k) k090140.(b)(4).Summary of investigational findings: the used device is returned in original packaging.The filter loaded on the femoral introducer.The secondary filter legs/wires, normally shaped as leaves, are deformed.Furthermore, marks are found on the introducer sheath tip.Both observations strongly indicate, that filter has been advanced out of sheath and afterwards attempted to be retracted into sheath, causing marks on the sheath tip and deformed secondary filter wires/leaves.No evidence to suggest that device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to initial reporter: initially, ivc filter system went inside the sheath smoothly.But when the physician tried to reposition and retrieve the filter within the sheath, he was not possible to do so and resulted in deformation of the filter.Hence, the physician withdrew the defective filter and used another ivc filter system.Patient outcome: unknown as information was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP JUGULAR & FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
4632 bjaeverskov
DA 
Manufacturer Contact
marianne hoey
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key6275485
MDR Text Key65665993
Report Number3002808486-2017-00068
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-UNI
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/19/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52
-
-