| Model Number |
VAMC4036C150TU |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/09/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A valiant captivia stent graft system was implanted in a patient for the endovascular treatment of a 6 cm diameter thoracic aortic aneurysm.It was reported that the patient presented for a routine follow up.The ct revealed that the 44x44x100 valiant stent graft had a distal type i endoleak.The distal thoracic aorta had changed from 36 mm in diameter to 42 mm in diameter.The physician implanted a 40x36x150 valiant stent graft deploying it in the previously implanted proximal 40x40x100 valiant extending down in the native artery completely covering the 44x44x100 valiant.However, the distal type i endoleak did not resolve.The physician elected to monitor the patient.The physician stated that the cause of the distal type i endoleak was due to disease progression and distal aneurysm dilation.No clinical sequelae were reported and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information: sex, date of birth.If information is provided in the future, a supplemental report will be issued.
|
| |
|