• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS SYMFONY; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); No Code Available (3191)
Date of Event 12/22/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxr00 intraocular lens (iol) was explanted from a female patient's left eye due to visual disturbance, inadequate and blurry vision at near.There was no incision enlargement required.The lens was replaced with a zmb00 26.5 diopter iol.The patient has now recovered.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 3/8/2017.Device returned to manufacturer ¿ yes.Device evaluation the device was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in half, most probably to make explant possible.Considering the condition of the lens, additional analysis is not possible.The customer's reported event could not be confirmed in the returned lens sample.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed that no similar complaints were received for this production order number.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6278290
Report Number9614546-2017-00051
Device Sequence Number1075626
Product Code POE
UDI-Device Identifier05050474579217
UDI-Public(01)05050474579217(17)210927
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/27/2021
Device Model NumberZXR00
Device Catalogue NumberZXR00U0260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/04/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/26/2017
Supplement Report FDA Received Date05/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-