• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. FMP HIP; LINER, ACETABULAR, FMP CONSTRAINED 28XMP8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. FMP HIP; LINER, ACETABULAR, FMP CONSTRAINED 28XMP8 Back to Search Results
Catalog Number 436-28-008
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/04/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the constrained liner breaking and the patient having dislocated.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the constrained liner broke and the patient dislocated.The in-vivo length of patient service for the implant was 2.6 months.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to (b)(4) for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had an acetabular liner that broke and the patient subsequently dislocated.No information was supplied with the complaint describing the nature of the break or conditions under which the break occurred.The scope of this investigation is limited without having the parts available to (b)(4) for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP HIP
Common Device Name
LINER, ACETABULAR, FMP CONSTRAINED 28XMP8
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key6280648
Report Number1644408-2017-00038
Device Sequence Number11051501
Product Code KWZ
UDI-Device Identifier00888912085717
UDI-Public(01)00888912085717
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K023794
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/26/2021
Device Catalogue Number436-28-008
Device Lot Number500C1009
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/04/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/26/2017
Supplement Report FDA Received Date03/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
436-28-000,LOT 496C1061
Outcome Attributed to Adverse Event Required Intervention;
-
-