OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/26/2016
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to olympus medical systems corp.But was returned to olympus (b)(4)).(b)(4) sent the subject device to a third party laboratory for microbiological testing and the testing indicated no microorganisms growth for the subject device.The manufacturing record of the subject device was reviewed with no irregularity relating to the phenomenon.The exact cause of the event could not be concluded at this moment.If additional and significant information becomes available, this report will be supplemented.
 
Event Description
Olympus was informed that the subject device was tested positive when the user facility conducted a routine microbiological test.There was no patient infection reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
hiroki moriyama
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
8142642517
MDR Report Key6282294
MDR Text Key66226116
Report Number8010047-2017-00088
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PK072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-