| Model Number |
MT22719-BLU |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The g5 system is associated with product code pqf.
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(6) 2017, to claim intermittent audio output on (b)(6) 2017.There was no report of any injury or medical intervention.No additional event or patient information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and the reported fault could not be reproduced and there was no failure detected.A review of the downloaded receiver log did not find any errors related to the customer complaint.A manual drop test for intermittency was performed and the test passed.The reported event of an intermittent audio output was not confirmed.A root cause could not be determined.
|
| |
|