• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MANUFACTURING INC CLINICAL CHEMISTRY CARBON DIOXIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MANUFACTURING INC CLINICAL CHEMISTRY CARBON DIOXIDE Back to Search Results
Catalog Number 03L80-31
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/07/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This issue was previously reported under mdr number 1628664-2016-00298 under a different suspect device.The customer reported that some physicians are questioning low co2c results at one of their facilities.The customer did a study between different facilities and noticed the site with architect analyzer co2c results are 30% lower on all patients.The facility with (b)(4) analyzer showed only 5% lower result.A review of complaints found no trends or similar complaints identified for the list number.A review of nonconformances for the list number did not identify any issues related to this complaint.A review of the carbon dioxide reagent package insert and the architect system operations manual shows adequate labeling for the customer's issue is provided.Based on this investigation, neither a malfunction nor a deficiency was identified for the carbon dioxide reagent list number (b)(4).
 
Event or Problem Description
The customer stated that the architect analyzer is generating falsely decreased co2 results compared to their sister facility.The result example provided: sample from (b)(6) 2016 co2 =17 (range 20-29)/ redrawn on (b)(6) 2016 = 15 / physician questioned the result so patient was redrawn at sister site and the (b)(4) = 24 (range 21-31).There was no reported impact to patient management.There was no additional patient information provided.
 
Additional Manufacturer Narrative
The device evaluation was reassessed and concluded that a malfunction occurred, therefore the device was not performing as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINICAL CHEMISTRY CARBON DIOXIDE
Common Device Name
CARBON DIOXIDE
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
MDR Report Key6289436
Report Number1628664-2017-00019
Device Sequence Number1564152
Product Code KHS
UDI-Device Identifier00380740005962
UDI-Public00380740005962
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K060295
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03L80-31
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/10/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/31/2017
Supplement Report FDA Received Date04/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT C16000 SYSTEM, LIST # 03L77-01,; SERIAL #(B)(4)
-
-