| Model Number |
25058 |
| Health Effect - Clinical Code |
Failure to Anastomose (1028)
|
| Date of Event |
01/09/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
We are in the process of performing the investigation and will submit the follow-up report once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
Report received stated the flixene graft was placed for dialysis in a patient but the anastomosis was not done correctly.The graft was cut too short and kinked.Patient was given a catheter for dialysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint sample was not returned for evaluation.The investigation revealed that the surgeon had inadvertently cut the graft too short during the procedure, therefore the anastomosis was not done correctly.The instructions for use describe the importance of graft length in the precautions section."grafts must be fully extended prior to sizing length for implantation." summary/conclusion: the graft was not retuned for evaluation.The device history records for this lot of grafts was reviewed and all of the test data met specifications.Based upon the additional event information provided by the facility it was concluded that during the procedure the surgeon inadvertently cut the graft short which caused the patient complication.
|
| |
|