• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; SCREW Back to Search Results
Model Number 25-878-05-91
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/03/2016
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor was repairing an infraorbital rim fracture in the patient.After drilling a pilot hole, she then proceeded to place the screw into the predrilled hole.The screw snapped off slightly before the screw head reached the plate.The body of the screw remains in the patient.No reported injuries occurred.
 
Additional Manufacturer Narrative
An investigation was performed using visual and stereo microscope inspection of the screw head returned.Tensile cracks were observed under the stereo microscope.Further observation determined there were no indications of material or manufacturing defects.The device history records could not be reviewed since no lot number was provided by the reporter.The results of the investigation conclude that the root cause for the breakages were due to mechanical overload on the device.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
MDR Report Key6291327
Report Number9610905-2017-00001
Device Sequence Number13674528
Product Code HRS
UDI-Device Identifier00888118047694
UDI-Public(01)00888118047694
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K971297
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number25-878-05-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/03/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/31/2017
Supplement Report FDA Received Date06/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age48 YR
-
-