WILLIAM COOK EUROPE GÜNTHER TULIP VENA CAVA FILTER JUGULAR APPROACH; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-JP-JUG-TULIP
Device Problems Kinked (1339); Failure to Advance (2524)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/08/2013
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Catalog #: igtcfs-65-jp-jug-tulip.Similar to device under 510(k) k090140.(b)(4).Summary of investigational findings: three minor kink appears on the sheath approx.17 cm from the fitting.Furthermore, two severe kink appears on the sheath respectively 19.5 cm and 22.5 cm from the fitting.The kink which are placed 19.5 cm from the fitting has a mark, probably added to the sheath by a primary filter leg.However, there are no perforation of the sheath.Under normal conditions the sheath is strong enough to accomplish the procedure, but it has been seen before that the sheath may kink if somehow exposed to excessive force because of tortuous anatomy.According to the description of the event resistance was met when introducer was advanced through the sheath but procedure was abandoned.No evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to initial reporter: the patient was an emergency outpatient.The sheath was advanced to the ivc.Right after the filter introduer was advanced into the sheath, resistance was encountered, and the resistance kept the filter introducer from advancing despite several attempts.Then, the whole delivery system was removed from the patient, and another filter was placed.Patient outcome: the patient is hospitalized.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GÜNTHER TULIP VENA CAVA FILTER JUGULAR APPROACH
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
4632 bjaeverskov
DA 
Manufacturer Contact
marianne hoey
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key6292166
MDR Text Key66216205
Report Number3002808486-2017-00241
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00827002037731
UDI-Public(01)00827002037731(17)160201(10)E3041313
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Physician
Type of Report Initial
Report Date 07/09/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-JP-JUG-TULIP
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/09/2013
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
-
-