Catalog Number 0036550
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code Not Applicable (3189)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One (1) 0036550 1/4" x 12' suction tubing package was returned to conmed for evaluation of "insufficient heatseal." the packaging test results revealed an open seal with adhesive transfer less than 1/4 inch, therefore this complaint is confirmed.The device was manufactured 1-july-2016.A review of the device history record for this lot found no discrepancies during the manufacturing process and no non-conformances regarding this product's identity, quality, safety, effectiveness or performance that could have caused or contributed to this event.Of the lot containing (b)(4) units there have been 2 complaints related to this reported complaint of "insufficient heatseal." a 2-year history of this device family revealed a total of (b)(4) complaints involving (b)(4) devices of which (b)(4) have been confirmed or pending evaluation for this failure mode and product.During this same 2-year time frame, over (b)(4) units were sold making this failure rate (b)(4) percent.This failure mode is addressed in the fmea and the safety risk has been found to be acceptable.The reported packaging anomalies were obvious to the distributor and therefore prompted the return of the devices for evaluation and replacement.To date, there have been no serious injuries or death related to this reported problem.However, to prevent future recurrences, an investigation has been initiated to address this issue.
 
Event or Problem Description
The distributor in (b)(4) reported that during receiving and inspection of incoming products, 1package containing 1/4" x 12' suction tubing was discovered with "insufficient heat seal." there was no patient involvement with this reported problem, as the packaging anomaly was discovered during inspection at the distributor facility prior to distribution to the end-user.This report is being filed based on the potential for injury with recurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1/4" X 12' SURGICAL TUBING
Common Device Name
SUCTION TUBING
Manufacturer (Section D)
CONMED CORPORATION
525 french road
utica NY 13502 5994
MDR Report Key6294300
Report Number3007305485-2017-00021
Device Sequence Number925176
Product Code BYY
UDI-Device Identifier00653405000528
UDI-Public(01)00653405000528
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P/A EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 01/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date06/30/2021
Device Catalogue Number0036550
Device Lot Number201607014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2017
Initial Report FDA Received Date02/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-