• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
Date of Event 01/05/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If implanted; give date: na (not applicable) lens was removed/replaced during the same procedure.If explanted; give date: na (not applicable) lens was removed/replaced during the same procedure.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was removed and replaced during the initial procedure.During phaco, the capsule broke.This affected the patent's left eye.The implanted symphony lens was not to blame.The lens was inserted, the incision enlarged, vitrectomy performed, and sutures were used.There was no post-op patient injury reported.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 03/08/2017.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection showed that the lens was damaged, most probably to make the explant possible.The customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6295407
Report Number9614546-2017-00063
Device Sequence Number1056457
Product Code POE
UDI-Device Identifier05050474605770
UDI-Public(01)05050474605770(17)211026
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date10/26/2021
Device Model NumberZXT225
Device Catalogue NumberZXT225U190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/11/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/02/2017
Supplement Report FDA Received Date05/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-