• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - SINGAPORE HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SINGAPORE HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C4479
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a small hole in a cassette when it was removed for re-priming.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Underwater pressure leak testing was performed with no issues noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
singapore
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6298403
MDR Text Key66451843
Report Number1416980-2017-00851
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberR5C4479
Device Lot NumberS16E13087
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/13/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
HOMECHOICE
-
-