• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE GÜNTHER TULIP VENA CAVA FILTER JUGULAR APPROACH; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GÜNTHER TULIP VENA CAVA FILTER JUGULAR APPROACH; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-JP-JUG-TULIP
Device Problems Kinked (1339); Failure to Advance (2524)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/17/2013
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Catalog#: igtcfs-65-jp-jug-tulip.Similar to device under 510(k): k090140.(b)(4).Summary of investigational findings: the description of event states that "there was tortuosity in patient's anatomy".Investigation of the introducer sheath demonstrates two kinks at 13 and 27 cm, the last kink represents the kink where the filter has perforated the sheath.Based on the information provided, it is not possible to conclude an exact root cause for the filter perforating the sheath.However it is considered likely that the filter was attempted advanced through a kinked sheath.In normal conditions the sheath is strong enough to accomplish the procedure, but it has previously been seen that the sheath may kink if somehow exposed to excessive force due to tortuous anatomy.If filter is advanced through a kinked sheath, the filter may exceed the sheath wall.No evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to initial reporter: the device was used for ivc filter placement by right jugular approach.There was tortuosity in patient's anatomy.After the sheath was placed, the filter was advanced through the sheath.Then strong resistance was met and it was florospically confirmed that filter legs had penetrated the sheath.The penetrated filter legs were put back inside the sheath and the whole system was removed together with the sheath.The device was replaced with another one and the procedure was completed with no problems.Patient outcome: no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GÜNTHER TULIP VENA CAVA FILTER JUGULAR APPROACH
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
4632 bjaeverskov
DA 
Manufacturer Contact
marianne hoey
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key6299512
MDR Text Key66519149
Report Number3002808486-2017-00261
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00827002037731
UDI-Public(01)00827002037731(17)150329(10)E2896728
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Physician
Type of Report Initial
Report Date 12/18/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-JP-JUG-TULIP
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2013
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-