| Model Number |
WEP-4208A |
| Medical Device Problem Codes |
Radiofrequency Interference (RFI) (2314); Device Displays Incorrect Message (2591); Device Operational Issue (2914); Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/05/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer reported that the wep telemetry system signal showed a zig zag waveform on multiple channels.The customer sent the unit in for evaluation and repair and the evaluation has not yet been conducted.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the wep telemetry system signal showed a zig zag waveform on multiple channels.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported that the wep telemetry system signal showed a zig zag waveform on multiple channels.The customer sent the wep telemetry system unit in for evaluation and repair.The unit was cleaned and evaluated.The reported problem could not be duplicated through testing, troubleshooting and extended operation of the unit.Review of the device history indicates no previous cnd status.The unit was tested per the operator's manual and the results were recorded on the maintenance check sheet.The unit operates to manufacturer's specifications.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Search Alerts/Recalls
|