• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2R
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The scope was returned to olympus for evaluation.The evaluation confirmed the reported device damage.The bending section of the scope was found cracked with the skeleton ribs protruding through the bending section rubber near the insertion tube.There was also a hole and leak noted on the bending section rubber.The bending section rubber adhesive was also found chipped.The scope failed leak test.The scope was serviced and returned to the user facility.This type of bending section damage is attributed to excessive force.The instruction manual for use states, "do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.¿.
 
Event Description
Olympus was informed that the scope fibers were found exposed through the bending section.The scope was not used in a procedure at the time the device damage was noted.No patient or user injury reported.
 
Manufacturer Narrative
This supplemental report is being submitted to make a correction on the procode from nwb to fgb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
connie tubera
2400 ringwood avenue
san jose, CA 95131
408935-512
MDR Report Key6301218
MDR Text Key66549325
Report Number2951238-2017-00056
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2R
Device Catalogue NumberURF-V2R
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/20/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-