It was reported that the customer passed away at home, in bed.The caller stated autopsy will be performed; presumed cause of death is cardiac arrest.The customer had neuropathy, vascular disease, heart disease, and infections.The caller stated that the customer suffered from a lot of low blood glucose events.The customer's last recorded blood glucose value was 119 mg/dl on (b)(6) 2017 at 9:30pm, in the bolus history, with 5.7 units of bolus for 46 carbs.The customer was wearing the insulin pump at the time of death.The customer was not using sensors.The caller agreed returning the insulin pump for analysis.
|
The insulin pump was received with (b)(6) alkaline battery.The insulin pump passed all functional testing including idle current measurement test, run current measurement test, self-test, off no power test, unexpected restart error test, displacement test, basic occlusion test, occlusion test, prime test, rewind test and excessive no delivery test.The insulin pump passed the displacement accuracy test.However, the insulin pump was received with intermittent act and up button response; unable to determine root cause due to insulin pump preservation.The insulin pump had cracked reservoir tube lip and minor scratched lcd window.Data analysis: there is no data listed for the dated of (b)(6) 2017 due to the insulin pump was received with a depleted battery installed.
|