| Model Number |
ZXT150 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); Halo (2227); No Code Available (3191)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, was not provided.The lens remains implanted.(b)(4).Device evaluation the device was not returned to the manufacturer as the lens remains implanted.Device inspection could not be performed, therefore the reported complaint could not be verified.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this order number to date.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
Initially, it was reported that the patient was not happy with distant vision.After the implantation of the symfony lenses, patient experienced halos and starburst on the sides, in all lights.Additional information indicated that the physician is evaluating patient symptoms and hopes that the solution will simply be a lasik touch up for the minor refractive error on both eyes.Further information reported that no explant at this time, patient is wearing corrective lenses for fine tuning distance vision.However, patient will probably need lasik touch.The physician reported patient was at 3 weeks post op on (b)(6) 2016, and feels better with darker glasses.Patient does not have any corneal changes or dry eyes.Patient has -0.75d of residual astigmatism in both eyes.Furthermore, the physician indicated that the lens has a cloudy central area.Doctor was questioning if the lathe lines that he saw at the central, 0.75mm of the symfony lens could cause myopia.Reportedly, at this time, patient did not want any intervention performed and is content with wearing glasses.This report will capture zxt150 lens implanted on the right eye.A separate mdr report will be filed for the lens implanted on the left eye.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: photos of the lens implanted were reviewed.Better quality, slit beam photos need to be obtained to evaluate any opacity and location.Based upon the photos received, it was not possible to define a product deficiency as it is difficult to make anything of the photos as photo quality was not optimum.All pertinent information available to the manufacturer has been submitted.
|
| |
|
Search Alerts/Recalls
|