| Model Number |
B2 70071-P |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
01/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The failed iv set is being sent to the contract supplier for evaluation.A quality alert has been sent to the contract supplier on 01/18/2017 regarding this issue.
|
| |
|
Event or Problem Description
|
|
The user experienced a pinch clamp leakage problem.This is a ltc (long term care) pharmacy.The issue was discovered at pharmacy during routing testing.During routine pump testing, 2 tube sets showed evidence of leaking through closed clamps-one while threaded in a pump (top clamp only), one while hanging attached to liter bag (both upper & lower roller clamps engaged).Over the course of several hours, all fluid in both source bags slowly leaked past the clamps and the bags emptied.This was after the pump test was completed but the tubing was still attached to source bags.Test was with 1000ml bags of ½ ns (normal saline).No patient was involved in the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Zyno medical has received the investigation report from the contract manufacturer on 04/25/2017.The user-reported complaint was not confirmed.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795-2017-00027).
|
| |
|