STRYKER ORTHOPAEDICS-MAHWAH SIZE 8 ACCOLADE II 132 DEG; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6720-0837
Medical Device Problem Codes Device Contaminated During Manufacture or Shipping (2969); Manufacturing, Packaging or Shipping Problem (2975); Material Integrity Problem (2978)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Date of Event 01/11/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported on the implant, looks like frizzy white cotton like material.It was reported they decided not to implant and allegedly kept implant and original box.
 
Additional Manufacturer Narrative
An event regarding foreign material involving a accolade stem was reported.The event was confirmed.Method and results: -device evaluation and results: a visual inspection of the returned device was carried out.This noted a fibre like material on the ha coating of the device.The device was otherwise unremarkable.-medical records received and evaluation: no medical records or x-rays were made available for evaluation.-device history review: review indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: review indicated there have been no other events for the reported lot.Conclusions: (b)(4) has been raised to investigate the event.(b)(4) was raised on 23-feb-2017 and closed on 31-mar-2017.The scope and corrective actions are detailed within the nc and corresponding (b)(4).(b)(4) determined the fiber was most likely introduced to the accolade ii stem in the ha coating and packaging business units.
 
Event or Problem Description
It was reported on the implant, looks like frizzy white cotton like material.It was reported they decided not to implant and allegedly kept implant and original box.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIZE 8 ACCOLADE II 132 DEG
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key6307144
Report Number0002249697-2017-00475
Device Sequence Number768896
Product Code LPH
UDI-Device Identifier04546540664518
UDI-Public(01)04546540664518(11)160222(17)210228(10)54727401
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number6720-0837
Device Lot Number54727401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/11/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/07/2017
Supplement Report FDA Received Date04/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-