| Catalog Number |
6720-0837 |
| Medical Device Problem Codes |
Device Contaminated During Manufacture or Shipping (2969); Manufacturing, Packaging or Shipping Problem (2975); Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
| Date of Event |
01/11/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported on the implant, looks like frizzy white cotton like material.It was reported they decided not to implant and allegedly kept implant and original box.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding foreign material involving a accolade stem was reported.The event was confirmed.Method and results: -device evaluation and results: a visual inspection of the returned device was carried out.This noted a fibre like material on the ha coating of the device.The device was otherwise unremarkable.-medical records received and evaluation: no medical records or x-rays were made available for evaluation.-device history review: review indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: review indicated there have been no other events for the reported lot.Conclusions: (b)(4) has been raised to investigate the event.(b)(4) was raised on 23-feb-2017 and closed on 31-mar-2017.The scope and corrective actions are detailed within the nc and corresponding (b)(4).(b)(4) determined the fiber was most likely introduced to the accolade ii stem in the ha coating and packaging business units.
|
| |
|
Event or Problem Description
|
|
It was reported on the implant, looks like frizzy white cotton like material.It was reported they decided not to implant and allegedly kept implant and original box.
|
| |
|