BARD ACCESS SYSTEMS POWERGROSHONG 5F SL SILICONE CATHETER WITH MICROINTRODUCER BASIC TRAY (45CM); PERCUTANEOUS, IMPLANTED, LONG TERM INTRAVASCULAR CATHETER
|
Back to Search Results |
|
| Model Number |
N/A |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation, as the device was discarded after the event occurred.A lot history review (lhr) of rear0648 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the picc was placed on (b)(6) 2016, for chemotherapy.The patient was seen in the clinic on (b)(6) 2017, to disconnect the infusor bottle.It was stated that while flushing there was leaking at the site.The picc was removed in the clinic by the hcn; it was alleged that there was a crack at approximately 2cms below the external marking.Patient stated that they had not adhered to the activity restrictions at all times while the picc was in place and continued to remain active while receiving treatment, ¿possibly leading to fatigue of picc at entry to vessel.¿.
|
| |
|
Search Alerts/Recalls
|
|
|