Model Number 420676
Medical Device Problem Codes Material Disintegration (1177); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Based on the available information, this event is deemed to be a reportable malfunction.Additional information has been requested, however to date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event or Problem Description
It was reported by a nurse that the dressing disintegrated after seven days of use on the patients right leg/foot area.Photos were included.No further details have been provided.
 
Additional Manufacturer Narrative
Updated data: brand name: aquacel ag extra.Batch records have been reviewed; all required specification was met and documented within the batch records.All preventative maintenance was completed on this manufacturing line.Machine logs were reviewed and there were no issues reported relating to the complaint issue.(4) photographs were received and these confirm the dressings have disintegrated.A previous complaint issue was opened to investigate this issue which has now been completed and closed.It was noted that: ¿this appearance is not uncommon and it suggests that the wound was heavily exuding so that the hydrofiber gelled completely (as it¿s designed to do).The residue will most likely be on the dressing or the bandage (if no secondary dressing).From our perspective, there are no concerns but it would be good to confirm that the dressing met its specification.¿ a review of all process testing requirements within the batch history retention files identify that all are within the specification parameters.There were no discrepancies noted in the batch records related to the reported complaint issue.Therefore, this identifies that the product was manufactured per the quality system, validated processes and approved procedures in place at the time of manufacturing and packaging.This complaint issue is currently an isolated incident as no further like-for-like complaints have been received.This issue will be monitored through the post market product monitoring review process.No additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUACEL AG EXTRA
Common Device Name
DRESSING, WOUND, DRUG
MDR Report Key6308849
Report Number1000317571-2017-00007
Device Sequence Number19122548
Product Code FRO
UDI-Device Identifier00768455127375
UDI-Public768455127375
Combination Product (Y/N)N
PMA/510(K) Number
K121275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Nurse
Device Expiration Date07/31/2021
Device Model Number420676
Device Catalogue Number420676
Device Lot Number6F03578
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 01/16/2017
Supplement Date Received by Manufacturer12/14/2017
Initial Report FDA Received Date02/07/2017
Supplement Report FDA Received Date01/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-