| Catalog Number |
03.812.308 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: reportedly, there was no patient involvement.Date of event is unknown.Device is an instrument and is not implanted/explanted.Manufacturing location: (b)(4).Manufacturing date: 21.July 2015.No nonconformances were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a t-handle t25 stardrive shaft was stripped at the tip, the sleeve threads of a holding sleeve-standard with stop for matrix were stripped, the tip of the t-pal trial spacer has broken off, the tip of the t-pal spacer applicator has broken off, and the slap hammer broke into (2) pieces at the very lowest part of the hammer.This was discovered during routine inspection.There is no patient or case involved.This is report 2 of 5 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product investigation was completed.A visual inspection, device history record (dhr) review, and drawing review were performed as part of this investigation.The proximal coupling of the t-pal instruments had broken off.The broken fragment was not returned.The returned t-pal applicator inner shaft and trial spacer are utilized in the t-pal (transforaminal posterior atraumatic lumbar) system.After trialing, the applicator inner shaft is installed into the applicator handle and knob assembly until the release button clicks into place.The appropriate spacer is attached by rotating the knob clockwise until the security ring clicks into position displaying a green band.The knob can then continue to be rotated clockwise until tight; the implant will not pivot in this position.The implant can then be advanced into the intervertebral disc space with light hammering.Once in position the applicator knob is rotated counterclockwise until it stops at the security ring, allowing the implant to pivot.The implant can be advanced into the final position with light controlled hammering.Once the implant is in position, it can be detached by pushing the security ring down and simultaneously turning the applicator knob counterclockwise until it stops; the applicator can then be detached from the implanted spacer.Relevant drawings for the returned instrument were reviewed (both from the time of manufacture and present revision.The design, materials and finishing processes were found to be appropriate for the intended use of these devices.The received failure mode is consistent with impaction while the handle (03.812.001) security ring is pressed forward; this would concentrate impaction forces on the proximal coupling ball tip.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
This is report 2 of 6 for (b)(4).
|
| |
|