• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/18/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure, air was observed coming through the valve of the sheath. The sheath was replaced and the case was completed with cryo. No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the sheath was returned and visually inspected and functionally tested. Visual inspection of the sheath showed the device was intact with no apparent issues. Air aspiration was reproduced when a test balloon catheter was introduced through the sheath. Dissection showed the hemostatic valve was leaking. A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure, air was observed coming through the valve of the sheath. The sheath was replaced and the case was completed with cryo. No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFLEXCATH ADVANCE STEERABLE SHEATH
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6309731
MDR Text Key105637912
Report Number3002648230-2017-00067
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 03/07/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/09/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number84933
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-